Invacare Platinum IRC5LXO2AWQ Användarmanual

Typ
Användarmanual

Denna manual är också lämplig för

Dealer: This manual must be given to the end
user.
User: Before using this concentrator, read
this manual and save for future reference.
User Manual
Platinum
Oxygen Concentrators Standard,
with
/SensO
2
®
and HomeFill
®
Model IRC5LXAW
Model IRC5LXO2AW
Model IRC5LXO2AWQ
Model IRC9LXO2AWQ
0434
English UK . . . . . . . . . . . . . 1
Danish. . . . . . . . . . . . . . . . 29
Swedish. . . . . . . . . . . . . . . 58
Norwegian . . . . . . . . . . . . 87
Finnish. . . . . . . . . . . . . . . 119
Platinum
2 Part No 1141493
WARNING
DO NOT USE THIS PRODUCT OR ANY AVAILABLE
OPTIONAL EQUIPMENT WITHOUT FIRST
COMPLETELY READING AND UNDERSTANDING
THESE INSTRUCTIONS AND ANY ADDITIONAL
INSTRUCTIONAL MATERIAL SUCH AS OWNER’S
MANUALS, SERVICE MANUALS OR INSTRUCTION
SHEETS SUPPLIED WITH THIS PRODUCT OR
OPTIONAL EQUIPMENT. IF YOU ARE UNABLE TO
UNDERSTAND THE WARNINGS, CAUTIONS OR
INSTRUCTIONS, CONTACT A HEALTHCARE
PROFESSIONAL, DEALER OR TECHNICAL
PERSONNEL BEFORE ATTEMPTING TO USE THIS
EQUIPMENT - OTHERWISE, INJURY OR DAMAGE
MAY OCCUR.
ACCESSORIES
There are many different types of humidifiers, oxygen tubing,
cannulas and masks that can be used with this device. You
should contact your local home care provider for
recommendations on which of these devices will be best for
you. They should also give you advice on the proper usage,
maintenance, and cleaning.
The supply accessories (nasal cannula, mask, tubing,
humidifier, etc.) used to deliver oxygen to the patient need to
include a means to reduce the propagation of fire in the
accessories for the safety of the patient and others. If a
commercially available, fire-activated flow stop device is used
in the accessories setup, it should be placed as close to the
patient as practicable.
ACCESSORIES WARNING
Invacare products are specifically designed and
manufactured for use in conjunction with Invacare
accessories. Accessories designed by other manufacturers
have not been tested by Invacare and are not
recommended for use with Invacare products.
TABLE OF CONTENTS
Part No 1141493 3 Platinum
TABLE OF CONTENTS
SPECIAL NOTES................................................4
RECYCLING INFORMATION .............................. 5
SECTION 1 — GENERAL GUIDELINES.....................6
Maintenance.......................................................................................... 7
Radio Frequency Interference.......................................................... 7
To Reduce The Risk Of Burns, Electrocution, Fire Or Injury To
Persons .................................................................................................. 7
Oxygen Concentration...................................................................... 8
SECTION 2 — FEATURES.......................................9
SECTION 3 — SHIPPING AND HANDLING...........10
Unpacking............................................................................................10
Inspection............................................................................................10
Storage.................................................................................................10
SECTION 4 — TYPICAL PRODUCT PARAMETERS 12
SECTION 5 — OPERATING INSTRUCTIONS .........16
Intended Use ......................................................................................16
Introduction........................................................................................16
Select a Location ...............................................................................16
Recommended Guidelines for Optimum Performance............16
Plug in Power Cord ..........................................................................17
Connect Humidifier (If So Prescribed).........................................17
Power Switch .....................................................................................20
Flowrate...............................................................................................21
SensO
2
Oxygen Purity Indicators (LX O
2
Sensor Model).......22
Indicators (LX Model)......................................................................23
Main Power Loss Alarm System ....................................................23
Elapsed Time Meter (Hour Meter) ...............................................23
SECTION 6 — MAINTENANCE ............................24
Routine Maintenance........................................................................24
SECTION 7 — TROUBLESHOOTING GUIDE...........27
CUSTOMER SERVICE AND WARRANTY
INFORMATION............................................. 150
SPECIAL NOTES
Platinum
4 Part No 1141493
SPECIAL NOTES
Signal words are used in this manual and apply to hazards or
unsafe practices which could result in personal injury or property
damage. Refer to the following table for definitions of the signal
words.
NOTICE
The information contained in this document is subject to change
without notice.
DANGER
DO NOT SMOKE while using this device. Keep all
matches, lit cigarettes or other sources of ignition out of
the room in which this product is located and away from
where oxygen is being delivered.
NO SMOKING signs should be prominently displayed.
Textiles and other materials that normally would not
burn are easily ignited and burn with great intensity in
oxygen enriched air. Failure to observe this warning can
result in severe fire, property damage and cause physical
injury or death.
SIGNAL WORD MEANING
DANGER
Danger indicates an imminently hazardous situation
which, if not avoided, will result in death or serious
injury.
WARNING
Warning indicates a potentially hazardous situation
which, if not avoided, could result in death or
serious injury.
CAUTION
Caution indicates a potentially hazardous situation
which, if not avoided, may result in property
damage or minor injury or both.
RECYCLING INFORMATION
Part No 1141493 5 Platinum
RECYCLING INFORMATION
This product has been supplied from an environmentally aware
manufacturer that complies with the Waste Electrical and Electronic
Equipment (WEEE) Directive 2002/96/CE.
This product may contain substances that could be harmful to the
environment if disposed of in places (landfills) that are not
appropriate according to legislation.
The 'crossed out wheelie bin' symbol is placed on this product to
encourage you to recycle wherever possible.
Please be environmentally responsible and recycle this product
through your recycling facility at its end of life.
SECTION 1—GENERAL GUIDELINES
Platinum
6 Part No 1141493
SECTION 1—GENERAL GUIDELINES
WARNING
SECTION 1 - GENERAL GUIDELINES contains
important information for the safe operation and use of
this product.
In order to ensure the safe installation, assembly and operation of
the Platinum concentrator these instructions MUST be followed.
CAUTION
“Caution: Statutory law can restrict this device to sale by or on order
of a physician, or any other practitioner licensed by the law of the
governmental agency in which he/she practices to use or order the
use of this device.”
DANGER
Risk of electric shock. DO NOT disassemble. Refer servicing to
qualified service personnel.
WARNINGS
The use of oxygen therapy requires that special care be taken to
reduce the risk of fire. Any materials that will burn in air, and some
that will not, are easily ignited and burn rapidly in high
concentrations of oxygen. For safety concerns, it is necessary that all
sources of ignition be kept away from the product and preferably
out of the room in which it is being used. NO SMOKING signs
should be prominently displayed.
A spontaneous and violent ignition may occur if oil, grease or
greasy substances come in contact with oxygen under pressure.
These substances MUST be kept away from the oxygen
concentrator, tubing and connections, and all other oxygen
equipment. DO NOT use any lubricants unless expressly
recommended by Invacare.
For optimum performance, Invacare recommends that each
concentrator be on and running for a minimum of 30 minutes at a time.
Shorter periods of operation may reduce maximum product life.
If the concentrator has a damaged cord or plug, if it is not working
properly, if it has been dropped or damaged, or dropped into water,
call Qualified Service Personnel for examination and repair.
SECTION 1—GENERAL GUIDELINES
Part No 1141493 7 Platinum
WARNINGS
Keep the oxygen tubing, cord, and unit away from heated or hot
surfaces, including space heaters, blankets, stoves and similar
electrical appliances.
DO NOT move or relocate concentrator by pulling on
the cord.
NEVER drop or insert any object into any opening.
NEVER block the air openings of the product or place it on a
soft surface, such as a bed or couch, where the air opening may
be blocked. Keep the openings free from lint, hair and the like.
Fill humidifier with water to the level shown by the manufacturer.
DO NOT overfill.
Move oxygen concentrator at least 7.5 cm (3 in) away from walls,
draperies or furniture.
Invacare recommends that Crushproof oxygen tubing (supplied by
Invacare) be used with this product and NOT exceed 15.2 m (50 ft)
in length. Recommended use for Platinum 9 is up to 15.m (50 ft) in
length of high flow tubing.
Maintenance
The Invacare Oxygen Concentrator was specifically designed to
minimize routine preventive maintenance at intervals of once per
year. Only professionals of the healthcare field or persons fully
conversant with this process such as authorized or factory trained
personnel should perform preventive maintenance or performance
adjustments on the oxygen concentrator.
Radio Frequency Interference
Most electronic equipment is influenced by Radio Frequency
Interference (RFI). Caution should be exercised with regard to the
use of portable communications equipment in the area around such
equipment.
To Reduce The Risk Of Burns,
Electrocution, Fire Or Injury To Persons
Avoid using while bathing. If continuous usage is required by the
physician’s prescription:
The concentrator must be located in another room at least 2.1 m (7
ft) from the bath.
DO NOT come in contact with the concentrator while wet.
SECTION 1—GENERAL GUIDELINES
Platinum
8 Part No 1141493
WARNINGS
A product should NEVER be left unattended when plugged in.
DO NOT place or store product where it can drop into water or
other liquid.
DO NOT reach for product that has fallen into water. Unplug
IMMEDIATELY.
This device is to be used only in accordance with the prescription of
a physician and this Owners Manual. If at any time the patient or
attendant conclude that the patient is receiving an insufficient
amount of oxygen, the supplier and/or physician should be
contacted IMMEDIATELY. No adjustments should be made to the
flowrate unless prescribed by a physician.
Close supervision is necessary when this product is used near
children or physically-challenged individuals.
Use this product for only intended use as described in this manual.
DO NOT connect the concentrator in parallel or series with other
oxygen concentrators or oxygen therapy devices.
Use of some administration accessories or certain humidifiers not
specified for use with oxygen concentrator may impair the
performance.
In certain circumstances oxygen therapy can be hazardous.
Invacare recommends that you seek medical advice before using
this machine.
This equipment is to be used as an oxygen supplement and is not
considered life supporting or life sustaining.
Avoid creation of any spark near medical oxygen equipment. This
includes sparks from static electricity created by any type of friction.
Oxygen Concentration
Take care not to set the flow above RED ring. An oxygen flow
greater than 5 L/min (IRC5LXO2AW, IRC5LXAW,
IRC5LXO2AWQ) and 9 L/min (IRC9LXO2AWQ) will decrease the
oxygen concentration.
SECTION 2—FEATURES
Part No 1141493 9 Platinum
SECTION 2—FEATURES
Oxygen Purity and Fault Indicator
Lights (SensO
2
model)/Fault and
Power Indicator Lights (LX model)
Circuit Breaker
Power Switch
Elapsed Time Meter
(Hour Meter)
Cabinet Filter
(one of two
shown)
Flowmeter
Oxygen Outlet
FRONT VIEW
TOP VIEW
ACCESSORIES
(NOT SHOWN):
HomeFill home oxygen
compressor
Contact your local Invacare
representative for available
accessory information.
*NOTE: This outlet fitting is to be used only for filling oxygen
cylinders with the HomeFill home oxygen compressor. The outlet
fitting does not affect concentrator performance. Refer to the
HomeFill owner's manual, part number 1145806, for connection
and operating instructions. When not in use, the plug provided
with the concentrator should be inserted into the outlet fitting.
For more information about the HomeFill, contact Invacare.
Power Cord
Outlet Fitting*
SECTION 3—SHIPPING AND HANDLING
Platinum
10 Part No 1141493
SECTION 3—SHIPPING AND
HANDLING
NOTE: For this procedure, refer to FIGURE 3.1
Before you install and operate the concentrator, please refer to
Platinum Series Owner's Manual.
The Platinum concentrator should ALWAYS be kept in the upright
position to prevent cabinet damage while being transported. The
shipping container has been designed to assure maximum
protection of the concentrator.
If the concentrator is to be reshipped by common carrier, it should
be packed in a new carton. Additional cartons are available from
Invacare.
The air compressor suspension system has been engineered to
withstand severe motion and orientation.
Unpacking
1. Check for any obvious damage to the carton or its
contents. If damage is evident, notify the carrier, or
Invacare.
2. Remove all loose packing from the carton.
3. Carefully remove all the components from the carton.
NOTE: Retain carton and packing materials for storage of the Platinum.
Inspection
1. Examine exterior of the oxygen concentrator for nicks,
dents, scratches or other damages. Inspect all components.
Storage
1. Store the repackaged oxygen concentrator in a dry area.
2. DO NOT place other objects on top of the repackaged
concentrator.
SECTION 3—SHIPPING AND HANDLING
Part No 1141493 11 Platinum
FIGURE 3.1 Unpacking
SECTION 4—TYPICAL PRODUCT PARAMETERS
Platinum
12 Part No 1141493
SECTION 4—TYPICAL PRODUCT
PARAMETERS
Regulatory
Listing:
EN/IEC 60601-1
Alternating Current
Type BF equipment
Unit running
Unit not running
Read manual
DO NOT smoke
Class II, Double Insulation
No open flame
RISK OF ELECTRIC SHOCK
This product complies with Directive 93/42/EEC concerning medical
devices.
The launch date of this product is stated in the EC
declaration of conformity.
Electrical Requirements: 230 VAC ± 10% (253 VAC/207 VAC), 50 Hz
Rated Current Input: 1.4 A (IRC5LXO2AWQ)
2.0 A (IRC5LXAW, IRC5LXO2AW)
2.3 A (IRC9LXO2AWQ)
Sound Level: 39.5 dB Average (IRC5LXO2AWQ)
44 dB Average (IRC5LXAW, IRC5LXO2AW)
50 dB Average (IRC9LXO2AWQ)
Altitude: IRC5LXAW, IRC5LXO2AW
IRC5LXO2AWQ - Up to 1828 meters (6,000
ft.) above sea level without degradation of
concentration levels. NOT RECOMMENDED
FOR USE ABOVE 1828 meters (6,000 ft)
IRC9LXO2AWQ - Up to 4,000 ft (1230
meters) above sea level without degradation of
concentration levels.
SECTION 4—TYPICAL PRODUCT PARAMETERS
Part No 1141493 13 Platinum
*Oxygen Output
Concentration Levels:
All 5LXO2AWQ/5LXAW/
5LXO2AW/9LXO2AWQ
models
* (Stated concentration levels
achieved after initial warm-up
period (approximately 30
minutes).
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
93% minimum at 1 to 3 L/min
91% minimum at 4 L/min
87% minimum at 5 L/min (maximum
recommended flow)
IRC9LXO2AWQ
92% minimum at 1 to 2 L/min
91% minimum at 8 L/min
87% minimum at 9 L/min
93% minimum at 3 to 7 L/min
Maximum Outlet Pressure: IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
34.5 ± 3.45 kPa (5 ± 0.5 psi)
IRC9LXO2AWQ
62.0 kPa ± 3.45 kPa (9 ± 0.5 psi)
Flow Range: IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ 0.5 to 5 L/min Flowrates less
than 1 L/min are not recommend
IRC9LXO2AWQ
1 L/min to 9 L/min Flowrates less than 1 L/min.
are not recommended.
Potential Obstruction Alert: The concentrator detects a condition that may
indicate a potential obstruction of the output
oxygen.
Rapid audible beeping alert (this alert is
deactivated when accessories are connected).
May be associated with flow setting of 0.5
L/min or less.
Average Power Consumption: (IRC5LXO2AWQ)
300 W
280 W @ 3 L/min
(IRC5LXO2AW)
365 W
340 W @ 3 L/min
380 W (IRC5LXAW)
460 W (IRC9LXO2AWQ)
Pressure Relief Mechanism
Operational at:
241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)
Change in maximum
recommended flow when back
pressure of 7k Pa is applied:
0.7 L/min
Filters: Cabinet (2), Outlet HEPA and Compressor
Inlet.
>
>
SECTION 4—TYPICAL PRODUCT PARAMETERS
Platinum
14 Part No 1141493
Safety System: Current overload or line surge shutdown.
High temperature compressor shutdown.
High Pressure Alarm w/compressor
shutdown.
Low Pressure Alarm w/compressor
shutdown.
Battery Free Power Loss Alarm.
SensO
2
Oxygen System Possible Obstruction
Alert.
Width: 46.7 cm ± 1 cm (18 3/8 in ± 3/8 in)
Height: 67.0 cm ± 1 cm (6 3/8 in. ± 3/8 in)
Depth: 36.5 cm ± 1 cm (14 3/8 in. ± 3/8 in)
Weight: IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
23.6 kg ± 1 kg (52 lbs ± 2 lbs)
IRC9LXO2AWQ
24.0 kg ± 1 kg (53 lbs ± 2 lbs)
Shipping Weight: IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
26.8 kg ± 1 kg (59 lbs ± 2 lbs)
IRC9LXO2AWQ
27.2 kg ± 1 kg (60 lbs ± 2 lbs)
Operating Ambient
Temperature:
10°C - 35°C (50°F - 95°F) at 20-60% relative
humidity.
Exhaust: Less than Ambient +19°C (+ 35°F)
Oxygen Output: Less than Ambient +2°C (+ 4°F).
Cabinet: Impact Resistant flame-retardant plastic
cabinet that conforms to UL 94-V0.
Regulatory Listing: IRC5LXAW/IRC5LXO2AW/
IRC5LXO2AWQ/IRC9LXO2AWQ
ETL tested as complying with:
EN55011
EN61000-3-2
EN61000-3-3
IEC 60601-1, A1, A2
IEC 60601-1-2
IRC5LXO2AW/IRC5LXO2AWQ/
IRC9LXO2AWQ
Models Only
ISO8359
MDD
CE marked models: IRC5LXO2AW, IRC5LXO2AWQ,
IRC9LXO2AWQ
Electrical: No extension cords.
SECTION 4—TYPICAL PRODUCT PARAMETERS
Part No 1141493 15 Platinum
SENSO
2
INDICATORS
LX INDICATORS
Placement: No closer than 7.5 cm (3 inches) from any
wall, furniture, draperies, or similar surfaces.
Tubing: IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
2.1 m (7 ft) cannula with a maximum 15.2 m
(50 ft) of Crush-Proof Tubing (DO NOT
pinch).
IRC9LXO2AWQ
Recommended use up to 15.2 m (50 ft) high
flow tubing with high flow cannula at all flow
rates.
Relative Humidity: 20 to 60%
Time of Operation: Up to 24 hours per day.
Recommended Storage and
Shipping Temperature:
-29°C to 65°C (-20°F to 150°F) at 15-95%
relative humidity.
LABEL
SYMBOL
STATUS INDICATOR LIGHTS (LED)
SYSTEM OKAY
O
2
over 85%
GREEN Indicator Light
O
2
Between 73% to 85% YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
Call a qualified technician.
SYSTEM FAILURE
O
2
Below 73%
RED Indicator Light
Refer to Troubleshooting Guide
on page 27.
LABEL
SYMBOL
STATUS INDICATOR LIGHTS
(LED)
SYSTEM OKAY GREEN Indicator Light
SYSTEM FAILURE RED Indicator light
Continuous Audible Alarm
Sieve-GARD™ Compressor
Shutdown
Call a qualified technician
O
2
I/O
SECTION 5—OPERATING INSTRUCTIONS
Platinum
16 Part No 1141493
SECTION 5—OPERATING
INSTRUCTIONS
Intended Use
Your oxygen concentrator is intended for individual use by patients
with respiratory disorders who require supplemental oxygen. The
device is not intended to sustain or support life.
Introduction
Your oxygen concentrator is intended for individual use in the
home. It is an electronically operated device that separates oxygen
from room air. It provides high concentration of oxygen directly to
you through a nasal cannula. Clinical studies have documented that
oxygen concentrators are therapeutically equivalent to other types
of oxygen delivery systems.
Your provider will show you how to use your oxygen concentrator.
He/She should be contacted with any questions or problems
regarding your oxygen concentrator. This owners manual will tell
you about your concentrator and will serve as a reference as you use
your concentrator.
Select a Location
You may select a room in your house where using your oxygen
concentrator would be most convenient. Your concentrator can be
easily rolled from room to room on its casters.
Your oxygen concentrator will perform best when operated under
the following conditions. Usage in environments other than those
described may result in the need for increased equipment
maintenance. The air intake of the unit should be located in a well
ventilated area to avoid airborne pollutants and/or fumes.
Recommended Guidelines for Optimum
Performance
Temperature: (10°C - 35°C) 50°F - 95°F
Electrical: No extension cords.
Placement: No closer than 7.5 cm (3 in) from the wall,
furniture, draperies, or similar surfaces.
SECTION 5—OPERATING INSTRUCTIONS
Part No 1141493 17 Platinum
NOTE: Ensure that your concentrator is at least 7.5 cm (3 in) away
from walls, draperies or furniture to assure sufficient air flow. Avoid
deep pile carpets and heaters, radiators or hot air registers.
Plug in Power Cord
1. Plug in power cord to an electrical outlet.
Connect Humidifier (If So Prescribed)
NOTE: For this procedure, refer to FIGURE 5.1 on page 18, FIGURE
5.2 on page 18, FIGURE 5.3 on page 19 and FIGURE 5.4 on page 20.
1. Fill humidifier with distilled water to the level indicated by the
manufacturer. (Please contact Invacare for recommendations).
WARNING
DO NOT overfill humidifier.
Replace the humidifier cap and securely tighten.
Tubing and Cannula: IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
2.1 m (7 ft) cannula with a maximum 15.2 m
(50 ft) of Crush-Proof Tubing (DO NOT
pinch).
IRC9LXO2AWQ
Recommended use up to 15.2 m (50 ft) high
flow tubing with high flow cannula at all flow
rates.
Environment: Smoke and soot-free. No confined spaces
(Example: No closets).
Relative Humidity: 20 to 60%
Time of Operation: Up to 24 hours per day.
Flow Rate IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
0.5 L/min. to 5 L/min Flow rates less than 1
L/min., are not recommended.
IRC9LXO2AWQ
1 L/min. to 9L/min Flow rates less than 1
L/min are not recommended.
Minimum Operating Time: 30 Minutes
SECTION 5—OPERATING INSTRUCTIONS
Platinum
18 Part No 1141493
DO NOT reverse the oxygen input and output
connections. Water from the humidifier bottle will travel
through the cannula back to the patient.
FIGURE 5.1 Filling the Humidifier
2. Place the humidifier bottle in the humidifier compartment.
FIGURE 5.2 Humidifier Compartment
3. Remove the filter access door located on the side of the
concentrator. Insert a flathead screwdriver in the plate
groove on the top edge of the filter access door and gently
pry the filter access door open. Refer to Detail “A” in
FIGURE 5.3.
4. The humidifier adapter is next to the inlet filter. Pull up
and remove the humidifier bottle adapter. Refer to Detail
“B” in FIGURE 5.3.
5. Attach it to the humidifier by turning the wing nut on the
humidifier bottle counterclockwise until it is securely
fastened. Refer to Detail “C” in FIGURE 5.3.
Humidifier Bottle
Humidifier Bottle
SECTION 5—OPERATING INSTRUCTIONS
Part No 1141493 19 Platinum
FIGURE 5.3 Attaching the Humidifier Bottle Adapter
6. Attach oxygen tubing from the humidifier bottle to
oxygen outlet connector on the oxygen concentrator.
7. Attach the cannula/patient supply tubing to the
humidifier bottle outlet.
8. After assembly, ensure that oxygen is flowing through the
cannula.
9. Replace the filter access door on the side of concentrator
by snapping it back into the plate groove.
DETAIL “A”
Plate Groove (Insert Flat
Head Screwdriver Here)
Filter Access Door
DETAIL “B”
Humidifier Bottle
Adapter
Humidifier Bottle
Adapter
Filter
Access
Door
DETAIL “C”
SECTION 5—OPERATING INSTRUCTIONS
Platinum
20 Part No 1141493
FIGURE 5.4 Attaching the Oxygen Tubing
WARNING
For optimum performance, Invacare recommends that
each concentrator be on and running for a minimum of 30
minutes at a time. Shorter periods of operation may
reduce maximum product life.
Power Switch
NOTE: For this procedure, refer to FIGURE 5.5.
1. Press power switch to on
position. All the panel
lights and the audible
alarm will come on for
one second, indicating
that the unit is
functioning properly.
FIGURE 5.5 Power Switch
Oxygen
Tubing
Humidifier Outlet
Oxygen Outlet
Connector
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152

Invacare Platinum IRC5LXO2AWQ Användarmanual

Typ
Användarmanual
Denna manual är också lämplig för