Amprobe AM-ACDC1H Bruksanvisning

Kategori
Mätning
Typ
Bruksanvisning
ACDC-100 TRMS
ACDC-100
Versatile AC/DC Clamp-on
Multimeter Series
Users Manual
Mode d’emplois
Bedienungshandbuchs
Manuale d’Usos
Manual de usos
Användarhandboks
1
English
ACDC-100 TRMS
ACDC-100
Versatile AC/DC Clamp-on
Multimeter Series
Users Manual
ACDC100_Rev001
© 2008 Amprobe Test Tools.
All rights reserved.
2
Limited Warranty and Limitation of Liability
Your Amprobe product will be free from defects in material and workmanship for 1
year from the date of purchase. This warranty does not cover fuses, disposable batteries
or damage from accident, neglect, misuse, alteration, contamination, or abnormal
conditions of operation or handling. Resellers are not authorized to extend any other
warranty on Amprobe’s behalf. To obtain service during the warranty period, return the
product with proof of purchase to an authorized Amprobe Test Tools Service Center or
to an Amprobe dealer or distributor. See Repair Section for details. THIS WARRANTY
IS YOUR ONLY REMEDY. ALL OTHER WARRANTIES - WHETHER EXPRESS, IMPLIED OR
STAUTORY - INCLUDING IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE
OR MERCHANTABILITY, ARE HEREBY DISCLAIMED. MANUFACTURER SHALL NOT BE LIABLE
FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES,
ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the
exclusion or limitation of an implied warranty or of incidental or consequential damages,
this limitation of liability may not apply to you.
Repair
All test tools returned for warranty or non-warranty repair or for calibration should be
accompanied by the following: your name, company’s name, address, telephone number,
and proof of purchase. Additionally, please include a brief description of the problem or
the service requested and include the test leads with the meter. Non-warranty repair or
replacement charges should be remitted in the form of a check, a money order, credit card
with expiration date, or a purchase order made payable to Amprobe® Test Tools.
In-Warranty Repairs and Replacement – All Countries
Please read the warranty statement and check your battery before requesting repair.
During the warranty period any defective test tool can be returned to your Amprobe® Test
Tools distributor for an exchange for the same or like product. Please check the “Where to
Buy” section on www.amprobe.com for a list of distributors near you. Additionally, in the
United States and Canada In-Warranty repair and replacement units can also be sent to a
Amprobe® Test Tools Service Center (see address below).
Non-Warranty Repairs and Replacement – US and Canada
Non-warranty repairs in the United States and Canada should be sent to a Amprobe® Test
Tools Service Center. Call Amprobe® Test Tools or inquire at your point of purchase for
current repair and replacement rates.
In USA In Canada
Amprobe Test Tools Amprobe Test Tools
Everett, WA 98203 Mississauga, ON L4Z 1X9
Tel: 877-AMPROBE (267-7623) Tel: 905-890-7600
Non-Warranty Repairs and Replacement – Europe
European non-warranty units can be replaced by your Amprobe® Test Tools distributor for
a nominal charge. Please check the “Where to Buy” section on www.amprobe.com for a
list of distributors near you.
European Correspondence Address*
Amprobe® Test Tools Europe
In den Engematten 14
79286 Glottertal, Germany
Tel.: +49 (0) 7684 8009 - 0
*(Correspondence only – no repair or replacement available from this address. European
customers please contact your distributor.)
3
Hall-effect Clamp Jaw
for AC & DC current
electric field pick up
Hand/Finger Barrier
to indicate the limits
of safe access of
the meter during
measurement
Push-buttons for
special functions &
features
Input Jack for all
functions EXCEPT non-
invasive ACA & DCA
current functions
Common (Ground
reference) Input
Jack for all functions
EXCEPT non-invasive
ACA & DCA current
functions
Slide-switch Selector
to turn the power
ON/OFF and Select a
function
3-3/4 digits 4000
counts LCD display
Jaw trigger for
opening the clamp jaw
¡Jaw center Indicators,
at where best ACA
& DCA accuracy is
specified
4
ACDC-100 TRMS / ACDC-100
Versatile AC/DC Clamp-on Multimeter Series
CONTENTS
Symbols.............................................................................................................................................5
Safety Information...........................................................................................................................5
Unpacking and Contents .................................................................................................................6
Introduction......................................................................................................................................6
Features ...........................................................................................................................................6
Alignment marks .......................................................................................................................6
MAX O/ HOLD O .....................................................................................................................6
HOLD ..........................................................................................................................................7
MAX ............................................................................................................................................7
U Relative Zero mode ...............................................................................................................7
SELECT / Backlight ( ) ..............................................................................................................7
Manual or Auto-ranging ...........................................................................................................7
Auto Power Off function ..........................................................................................................7
Operation ........................................................................................................................................7
DC Voltage .................................................................................................................................7
AC Voltage .................................................................................................................................7
AC Current ..................................................................................................................................8
DC Current ..................................................................................................................................8
Resistance ...................................................................................................................................8
Continuity ...................................................................................................................................8
Diodes .........................................................................................................................................8
Capacitance ................................................................................................................................9
Maintenance and Repair ................................................................................................................9
Battery Replacement .................................................................................................................9
Specifications..................................................................................................................................10
General Specifications ..............................................................................................................10
Electrical ..................................................................................................................................11
5
SYMBOLS
1Battery :Refer to the manual
7Double insulated ;Dangerous Voltage
)Direct Current -Earth Ground
%Alternating Current 5Audible tone
Conforms to relevant Australian
standards. 3Complies with EU directives
Do not dispose of this product as
unsorted municipal waste. !Underwriters Laboratories.
[Note: Canadian and US.]
Application around and removal
from hazardous live conductors
is permitted
SAFETY INFORMATION
The ACDC-100 Series Digital Clampmeters conform to EN61010-1:2001; EN61010- 2-s
032:2002; CAT III 600 V, class 2 and pollution deg.2
This instrument is EN61010-1 certified for Installation Category III (600V). It is s
recommended for use in distribution level and fixed installations, as well as lesser
installations, and not for primary supply lines, overhead lines and cable systems.
Do not exceed the maximum overload limits per function (see specifications) nor the limits s
marked on the instrument itself. Never apply more than 600 Vdc/600 V ac rms between
the test lead and earth ground.
:Warnings and Precautions
Before and after hazardous voltage measurements, test the voltage function on a known s
source such as line voltage to determine proper meter functioning.
Disconnect the test leads from the test points before changing meter functions.s
Inspect the Clampmeter, test leads and accessories before every use. Do not use any s
damaged part.
Never ground yourself when taking measurements. Do not touch exposed circuit elements s
or test probe tips.
Do not operate the instrument in an explosive atmosphere.s
To reduce the risk of fire or electric shock, do not expose this product to rain or moisture.s
The meter is intended only for indoor use. To avoid electrical shock hazard, observe the s
proper safety precautions when working with voltages above 60 VDC or 30 VAC rms.
These voltage levels pose a potential shock hazard to the user.
6
Before and after hazardous voltage measurements, test the voltage function on a known s
source such as line voltage to determine proper meter functioning.
Keep your hands/fingers behind the hand/finger barriers (of the meter and the test leads) s
that indicate the limits of safe access of the hand-held part during measurement.
Inspect test leads, connectors, and probes for damaged insulation or exposed metal before s
using the instrument. If any defects are found, replace them immediately.
This Clamp-on meter is designed to apply around or remove from uninsulated hazardous s
live conductors. Individual protective equipment must be used if hazardous live parts of
the installation could be accessible.
Exercise extreme caution when: measuring voltage >20 V // current >10 mA // AC power s
line with inductive loads // AC power line during electrical storms // current, when the fuse
blows in a circuit with open circuit voltage >1000 V // servicing CRT equipment.
Remove test leads before opening the case to change the battery.s
Disconnect circuit power and discharge all high-voltage capacitors before testing s
resistance,continuity, diodes, or capacitance.
To avoid false readings, which could lead to possible electric shock or personal injury, s
replace the batteries as soon as the low battery indicator (0) appears.
UNPACKING AND CONTENTS
Your shipping carton should include
1 ACDC-100 or ACDC-100 TRMS
1 Test lead set
1 9V Battery (Installed)
1 Users Manual
1 Carrying Case
If any of the items are damaged or missing, immediately return the complete package to the
place of purchase for an exchange.
INTRODUCTION
The ACDC-100 and ACDC-100 TRMS Clamp-On meters are 1000 Amp / 600 V clamp meters
with 50 mm jaw opening for industrial applications. The features include large conductor
size and in-rush current measurement along with AC / DC voltage, AC / DC current, Resistance,
Continuity and Diode Tests.
FEATURES
Alignment marks (see Fig. 1)
Place conductor within the jaws at the intersection of the indicated marks as close as possible
to maximize the accuracy of the reading.
MAX O/ HOLD O
The HOLD feature freezes the display when the Obutton is pressed. The MAX feature
compares and displays the measured maximum value as fast as 30ms with auto-ranging
capability.
7
HOLD
Press the HOLD button momentarily toggles to hold mode for all of the functions. To release
the HOLD feature momentarily press the HOLD button.
MAX
Press the HOLD button for 1 second or more activates the MAX HOLD feature for the VDC,
VAC and ACA functions. To release the MAX HOLD feature press the HOLD button for 1
second or more.
U Relative Zero mode
Relative Zero mode allows the user to offset the subsequent meter measurements with
the displayed reading as the reference value. The display will now show readings relative
to the stored reference value. That is, display = reading - stored value. Pressing the button
momentarily toggles to relative mode.
SELECT / Backlight ( )
Press Backlight button more than 1 second, enable/disable Backlight.
Press the SELECT / Backlight button to step through the manually selected A&and 1
functions.
A)A A%and 1A5A*A(
Manual or Auto-ranging
Press the RANGE button momentarily to select manual-ranging mode. The LCD symbol
turns off. Press the button momentarily again to step through the Ranges. Press and hold the
RANGE button for more than 1 second to resume auto-ranging.
Auto Power Off function
The clamp meter powers down automatically after approximately 30 minutes of inactivity.
To turn it back on, move the function selector switch to OFF and back to a measuring
function.
To disable Auto Power Off, press and hold the HOLD button while moving the slide switch to
the desired function from OFF.
OPERATION
DC Voltage (see Fig. 1)
Set the Function Switch to 1. V).
Connect the test leads: Red to 2. +, Black to COM.
Connect the test probes to the circuit test points. 3.
Read the display, and if necessary, correct any overload (4. 0L) conditions.
AC Voltage (see Fig. 1)
Set the Function Switch to 1. V%.
Connect the test leads: Red to 2. +, Black to COM.
Connect the test probes to the circuit test points. 3.
Read the display, and if necessary, correct any overload (4. 0L) conditions.
8
AC Current (see Fig. 2)
Set the Function Switch to A1. &position.
Select AC current using the SELECT button. 2.
Open spring-loaded clamp by pressing the lever on left side of meter. 3.
Position clamp around one wire or conductor and release the clamp lever. Make sure that 4.
the clamp is entirely closed. The clamp must be positioned around only one conductor. If it
is placed around two or more current carrying conductors, the reading is FALSE.
Read the displayed value, and if necessary, correct any overload (5. 0L) conditions.
DC Current (see Fig. 2)
Set the Function Switch to A1. &position.
Select DC current using the SELECT button. 2.
Open spring-loaded clamp by pressing the lever on left side of meter. 3.
Position clamp around one wire or conductor and release the clamp lever. Make sure that 4.
the clamp is entirely closed. The clamp must be positioned around only one conductor. If it
is placed around two or more current carrying conductors, the reading is FALSE.
Read the displayed value, and if necessary, correct any overload (5. 0L) conditions.
CAUTION
Using the Resistance, Continuity, Diode or Capacitance functions on a live circuit will produce
false results and may damage the instrument. In many cases the suspected component must
be disconnected from the circuit to obtain an accurate measurement reading.
Resistance (see Fig. 3)
Set the Function Switch to 1. 1.
Connect the test leads: Red to 2. +, Black to COM.
Turn off power to the circuit being measured. 3.
Discharge any capacitors that may influence the reading. 4.
Connect the test probes across the resistance. 5.
Read the display. If6. 0L appears on the highest range, the resistance is too large to be
measured or the circuit is an open circuit.
Continuity (see Fig. 3)
Set the Function Switch to 1. 1 and press the SELECT button until 5is displayed.
Connect the test leads: Red to 2. +, Black to COM.
Turn off power to the circuit being measured. 3.
Discharge any capacitors that may influence the reading. 4.
Connect the test probes across the resistance or the two points of test. 5.
Listen for the tone that indicates continuity (>10 6. 1 and < 120 1).
Diodes (see Fig. 3)
Set the Function Switch to 1. 1 and press the SELECT button until *is displayed.
Connect the test leads: Red to 2. +, Black to COM.
9
Turn off power to the circuit being measured. 3.
Free at least one end of the diode from the circuit. 4.
Connect the test probes across the diode noting polarity. 5.
Read the display. A good diode has a forward voltage drop of about 0.6 V. An open or 6.
reverse biased diode will read .0L.
Capacitance (see Fig. 3)
Set the Function Switch to 1. 1 and press the SELECT button until (is displayed.
Connect the test leads: Red to +, Black to 2. COM.
Turn off power to the circuit being measured. 3.
Discharge the capacitor using a 100 k4. 1 resistor.
Free at least one end of the capacitor from the circuit. 5.
Connect the test probes across the capacitor. 6.
Read the display. 7.
Relative zero mode can be used to zero out the parasitic capacitance of the leads and the 8.
internal protection circuitry of the meter when measuring low capacitance in the order of
Pico Farad (pF).
MAINTENANCE AND REPAIR
If there appears to be a malfunction during the operation of the meter, the following steps
should be performed in order to isolate the cause of the problem:
Check the battery. 1.
Review the operating instructions for possible mistakes in operating procedure. 2.
Inspect and test the test leads for a broken or intermittent connection. 3.
Except for the replacement of the battery or test probes, repair of the multimeter should
be performed only by a Factory Authorized Service Center or by other qualified instrument
service personnel. The front panel and case can be cleaned with a mild solution of detergent
and water. Apply sparingly with a soft cloth and allow to dry completely before using. Do not
use aromatic hydrocarbons or chlorinated solvents for cleaning.
Battery Replacement (see Fig. 4)
:Warning
To prevent electrical shock or meter damage, disconnect the meter’s test leads from any
circuit and the meter, then turn the meter off before removing the battery cover. Battery
replacement should be performed in a clean environment and with appropriate care taken to
avoid contaminating the meter’s interior components.
Remove the screw and lift the battery cover. 1.
Replace the batteries with the same type (1.5V AAA). Note polarity guide below the 2.
battery.
Replace the battery cover and screw.3.
10
SPECIFICATIONS
General Specifications
Display: 3-3/4 digits 4000 counts LCD display
Update Rate: 3 per second nominal
Polarity: Automatic
Power Supply: standard 1.5V AAA Size (NEDA 24G or IEC R03) battery x
Power Consumption: typical 11 mA for ACA/DCA and 2.9 mA for other
functions
APO Timing: idle for 30 minutes
APO Consumption: typical 10µA for ACDC- 100 and 90µA for ACDC- 100
TRMS
Low battery: below approx. 2.4V
Environment: Indoor operation, below 2000 m
Temperature / Humidity:
Operating: 0°C to 40°C (14°F to 122 °F)
Relative humidity : < 80% RH @ 31°C decreasing linearly to 50% RH at 40°C
Storage: -20°C to 60°C (-4°F to 122 °F) / < 80% RH
Temperature Coefficient : 0. 5 x (specified accuracy)/ °C @ (0°C to 18°C or 28°C to
40°C)
Sensing : Average sensing for ACDC- 100; True RMS for ACDC- 100
TRMS
Clamp-on jaw: DC 1000A or AC 800A rms continuous
+ & COM terminals (all functions) : 600VDC/VAC rms
Jaw opening: 50mm max
Dimension : 227 x 78 x 40mm (8.9 x 3.1 x 1.6 in.)
Weight : 290 gm ( 0.6 lb)
Safety LVD: Meets EN60101-1:2001; EN61010-2-032(2002), Category II- 600 Volts ac
& dc; pollution degree : 2; class 2
3 EMC: This product complies with requirements of the following European Community
Directives: 89/336/EEC (Electromagnetic Compatibility) and 73/23/EEC (Low Voltage) as
amended by 93/68/EEC (CE Marking). However, electrical noise or intense electromagnetic
fields in the vicinity of the equipment may disturb the measurement circuit. Measuring
instruments will also respond to unwanted signals that may be present within the
measurement circuit. Users should exercise care and take appropriate precautions to avoid
misleading results when making measurements in the presence of electronic interference.
11
Electrical (23 °C ± 5 °C) < 75% RH
DC Voltage
RANGE Accuracy
400.0 mV ± (0.3% + 3dgts)
4.000V, 40.00V, 400.0V ± (0.5% + 3dgts)
600V ± (1.0% + 4dgts)
Resolution : 0.1 mV on 400 mV range
NMRR : >50dB @ 50/60Hz
CMRR : >20dB @ DC, 50/60Hz, Rs=1k1
Input Impedance : 10M1, 30pF nominal (1000M1 for 400.0mV range)
Max input : 600VDC/VAC rms
AC Voltage
RANGE Frequency Accuracy
400.0mV 1) 50Hz to 500Hz ± (4.0% + 4dgts)
4.000V, 40.00V, 400.0V 50Hz to 60Hz ± (1.0% + 4dgts)
4.000V, 40.00V, 400.0V 60Hz to 500Hz ± (1.5% + 4dgts)
600V 50Hz to 500Hz ± (2.0% + 4dgts)
Resolution: 0.1 mV on 400 mV range
CMRR : >60db @ DC to 60Hz, Rs= 1k1
Input impedance: 10M1, 30pf nominal
Crest factor (ACDC-100 TRMS): < 1.6: at full scale; < 3.2: at half scale
Max input: 600VDC/VAC rms
1) Selection by Range button manually, and is specified from AC 40mV (AC 60mV for ACDC-
100 TRMS) and up
Resistance
RANGE Accuracy
400.01± (0.8% + 6dgts)
4.000k1, 40.00k1, 400.0k1± (0.6% + 4dgts)
4.000M1± (1.0% + 4dgts)
40.00M1± (2.0% + 4dgts)
Resolution: 0.1 1 on 400 1 range
Open Circuit Voltage : 0.4VDC typical
Continuity Tester
Open Circuit Voltage: 0.4Vdc typical
Range: 400.01
Accuracy: ± (1.5% + 6dgts)
Audible threshold: between 101 and 1201.
Diode Tester
Open Circuit Voltage < 1.6 Vdc
Test Current (Typical) 0.4mA
12
Capacitance
RANGE 1) Accuracy 2) 3)
500.0 nf, 5.000 µf, 50.00 µf, 500.0 µf, 3000 µf ± (3.5% + 6dgts)
Resolution: 0.1 nf on 500.0 nf range
1) Additional 50.00nf range accuracy is not specified
2) Accuracies with film capacitor or better
3) Specified with battery voltage above 2.8V (approximately half full battery). Accuracy
decreases gradually to 12% at low battery warning voltage of approximately 2.4V
DC Current (Clamp-On)
RANGE Accuracy 1) 2)
0.0A to 400.0A ± (1.5% + 4dgts)
400A to 800A ± (1.5% + 4dgts)
800A to 900A ± (2.0% + 4dgts)
900A to 1000A ± (5.0% + 30dgts)
Max current: 1000A continuous
Resolution: 0.1 A on 400 A range
1) Induced error from adjacent current-carrying conductor: < 0.01A/A
2) Zero mode must be applied to zero the unit before measurement to ensure accuracy.
ACA Current (Clamp-on)
RANGE Frequency Accuracy 1) 2)
400.0A 5Hz to 40Hz ± (2.0% + 5dgts) 3)
400.0A 40Hz to 200Hz ± (1.5% + 5dgts)
< 50A 4) 200Hz to 400Hz ± (1.5% + 5dgts)
< 50A 4) 400Hz to 1kHz ± (2.0% + 5dgts)
400A to 600A 5Hz to 40Hz ± (2.0% + 5dgts) 3)
400A to 600A 40Hz to 100Hz ± (1.5% + 4dgts)
600A to 800A 5Hz to 60Hz ± (5.0% + 30dgts)
Max current: 800A rms continuous
Resolution: 0.1 A on 400 A range
1) Induced error from adjacent current-carrying conductor: < 0.01A/A
2) Crest Factor (ACDC-100 TRMS): < 1.6 : at full scale; < 3.2 : at half scale
3) ± (4.0% + 5dgts) for ACDC- 100 TRMS
4) Accuracy is specified at <50A in this frequency bandwidth due to limited calibrator output
capability for testing
13
Figure 1. — DC Voltage and AC Voltage functions
14
Figure 2. — ACA & DCA Current clamp-on function
15
Figure 3. — Resistance, Continuity, Capacitance and Diode functions
Figure 4. — Battery replacement
16
17
Français
ACDC-100 TRMS
ACDC-100
Pince multimètre C.A./C.C.
polyvalente
Mode d’emploi
ACDC100_Rev001
© 2008 Amprobe Test Tools.
Tous droits réservés.
18
Limites de garantie et de responsabilité
Amprobe garantit labsence de vices de mariaux et de fabrication de ce produit
pendant une période d’un (1) an prenant effet à la date d’achat. Cette garantie ne
s’applique pas aux fusibles, aux piles jetables ni à tout produit mal utilisé, modifié,
contaminé, négligé ou endommagé par accident ou soumis à des conditions anormales
d’utilisation et de manipulation. Les distributeurs agréés par Amprobe ne sont pas
autorisés à appliquer une garantie plus étendue au nom dAmprobe. Pour bécier
de la garantie, renvoyez le produit accompagné dun justificatif dachat auprès dun
centre de services agréé par Amprobe Test Tools ou d’un distributeur ou d’un revendeur
Amprobe. Voir la section Réparation pour tous les détails. LA PRESENTE GARANTIE EST
LE SEUL ET EXCLUSIF RECOURS. TOUTES AUTRES GARANTIES, EXPLICITES, IMPLICITES OU
STATUTAIRES, NOTAMMENT LE CAS ECHEANT LES GARANTIES DE QUALITE MARCHANDE
OU D’ADAPTATION A UN OBJECTIF PARTICULIER, SONT EXCLUES PAR LES PRESENTES. LE
FABRICANT NE SERA EN AUCUN CAS TENU RESPONSABLE DES DOMMAGES PARTICULIERS,
INDIRECTS, ACCIDENTELS OU CONSECUTIFS, NI DAUCUNS DEGATS OU PERTES DE DONNEES,
SUR UNE BASE CONTRACTUELLE, EXTRA-CONTRACTUELLE OU AUTRE. Etant donné que
certaines juridictions nadmettent pas les limitations dune condition de garantie implicite,
ou l’exclusion ou la limitation de déts accidentels ou consécutifs, il se peut que les
limitations et/ou les exclusions de cette garantie ne s’appliquent pas à votre cas.
Réparation
Tous les outils de test renvoyés pour une réparation ou un étalonnage couvert ou non par
la garantie doivent être accompagnés des éléments suivants : nom, raison sociale, adresse,
numéro de téphone et justicatif dachat. Ajoutez également une brève description
du problème ou du service demandé et incluez les cordons de mesure avec lappareil. Les
frais de remplacement ou de réparation hors garantie doivent être acquittés par cque,
mandat, carte de cdit avec date dexpiration ou par bon de commande payable à l’ordre
d’Amprobe® Test Tools.
Remplacements et réparations sous garantie – Tous pays
Veuillez lire la déclaration de garantie etrifier les piles avant de demander une
réparation. Pendant la période de garantie, tout outil de test défectueux peut être renvoyé
auprès de votre distributeur Amprobe® Test Tools pour être échangé contre un produit
identique ou similaire. Consultez la section « Where to Buy » sur le site www.amprobe.com
pour obtenir la liste des distributeurs dans votre région. Les appareils sous garantie devant
être remplacés ou réparés au Canada et aux Etats-Unis peuvent également être envoyés
dans un centre de services Amprobe® Test Tools (voir les adresses ci-dessous).
Remplacements et réparations hors garantie Canada et Etats-Unis
Les appareils à réparer hors garantie au Canada et aux Etats-Unis doivent être envos
dans un centre de services Amprobe® Test Tools. Appelez Amprobe® Test Tools ou
renseignez-vous auprès de votre lieu dachat pour connaître les tarifs en vigueur de
remplacement ou de réparation.
Aux Etats-Unis Au Canada
Amprobe Test Tools Amprobe Test Tools
Everett, WA 98203 E-U Mississauga, Ontario L4Z 1X9 Canada
Tél. : 877-AMPROBE (267-7623) Tél. : 905-890-7600
Remplacements et réparations hors garantie – Europe
Les appareils euroens non couverts par la garantie peuvent être remplacés par votre
distributeur Amprobe® Test Tools pour une somme nominale. Consultez la section
« Where to Buy » sur le site www.amprobe.com pour obtenir la liste des distributeurs
dans votre région.
Adresse postale européenne*
Amprobe® Test Tools Europe
In den Engematten 14
79286 Glottertal, Allemagne
Tél. : +49 (0) 7684 8009 - 0
*(servée à la correspondance – Aucun remplacement ou réparation nest possible à cette
adresse. Nos clients européens doivent contacter leur distributeur.)
19
Mâchoire de pince
ampèremétrique à
effet Hall pour les
mesures de courants
électriques c.a. et c.c.
sur le terrain
Collerette de
protection des doigts
indiquant les limites
d’accès sécurisé sur
l’appareil pendant
les mesures
Boutons poussoirs
des fonctionnalités
spéciales
Prise d’entrée pour
toutes les fonctions
SAUF les fonctions de
courant A c.a. et A c.c.
non invasives
Prise d’entrée
commune (référence
à la terre) pour
toutes les fonctions
SAUF les fonctions de
courant A c.a. et A c.c.
non invasives
Interrupteur
d’alimentation
marche/arrêt (ON/
OFF) et commutateur
de fonction
Ecran LCD à
3-3/4 chiffres,
4 000 comptes
Gâchette d’ouverture
des mâchoires de
la pince
¡ Indicateurs de
centrage des
mâchoires, là où la
meilleure précision
A c.a. et V. c.a.
est spécifiée
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98

Amprobe AM-ACDC1H Bruksanvisning

Kategori
Mätning
Typ
Bruksanvisning