GETINGE GROUP ARJOHUNTLEIGH ALPHA ACTIVE 4 Instructions For Use Manual

Typ
Instructions For Use Manual
ALPHA ACTIVE 4
Instructions for Use • Brugsvejledning
Instrucciones de uso • Mode d'emploi
Istruzioni per l'uso • Gebruiksaanwijzing • Bruksanvisning
English
Français
Dansk
Español
Italiano
Nederlands
Svensk
English
(i)
Contents
General Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
About Alpha Active 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Alpha Active 4 Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Alpha Active 4 Mattress Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Clinical Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Contra-Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Preparing the Alpha Active 4 system for use. . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Installing the Alpha Active 4 Mattress . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Controls, Alarms and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Alarms and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Quick Start. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Support Setting Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Changes in Patient Position. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
To Deflate and Store the Alpha Active 4 Mattress. . . . . . . . . . . . . . . . . . . . . . 10
CPR Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
To activate CPR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
To reset CPR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Decontamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Routine Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Alpha Active 4 System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Alpha Active 4 Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Alpha Active 4 Mattress Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Serial Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Technical Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Mattress. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Mattress Size Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Cover Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Cleaning Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
(ii)
(iii)
GENERAL SAFETY
Before you connect the system pump to a mains socket, read carefully all the installation
instructions contained within this manual.
The system has been designed to comply with regulatory safety standards including:
EN60601-1:1990/A13:1996 and IEC 60601-1:1988/A2:1995
UL60601-1 and CAN/CSA C22.2 No. 601.1-M90
Safety Warnings
It is the responsibility of the care giver to ensure that the user can use this product
safely.
Whilst the patient is unattended, safety sides should be used based on clinical
assessment and in line with local policy.
Alignment of the bed frame, safety sides and the mattress should leave no gap
wide enough to entrap a patient's head or body, or to allow egress to occur in a
hazardous manner where entanglement with the mains power cable and tubeset or
air hoses may result. Care should be exercised to prevent occurrence of gaps by
compression or movement of the mattress. Death or serious injury may occur.
Make sure that the mains power cable and tubeset or air hoses are positioned to
avoid causing a trip or other hazard, and are clear of moving bed mechanisms or
other possible entrapment areas. Where cable management flaps are provided
along the sides of the mattress, these should be used to cover the mains power
cable.
Electrical equipment may be hazardous if misused. There are no user-serviceable
parts inside the pump. The pump's case must only be removed by authorised
technical personnel. No modification of this equipment is allowed.
The mains power socket/plug must be accessible at all times. To disconnect the
pump completely from the electricity supply, remove the plug from the mains
power socket.
The CPR control and/or the CPR indicator tag must be visible and accessible at all
times.
Disconnect the pump from the mains power socket before cleaning and inspecting.
Keep the pump away from sources of liquids and do not immerse in water.
Do not use the pump in the presence of uncontained flammable liquids or gasses.
The cover of this product is vapour permeable but not air permeable and may
present a suffocation risk.
Only the pump and mattress combination as indicated by ArjoHuntleigh should be
used. The correct function of the product cannot be guaranteed if incorrect pump
and mattress combinations are used.
(iv)
Precautions
For your own safety and the safety of the equipment, always take the following
precautions:
Placing extra layers between the patient and the mattress potentially reduces the benefits
provided by the mattress and should be avoided or kept to a minimum. As part of sensible
pressure area care, it is advisable to avoid wearing clothing which may cause areas of
localised high pressure due to creases, seams, etc. Placing objects in pockets should be
avoided for the same reason.
Do not expose the system, especially the mattress, to naked flames, such as cigarettes,
etc.
Do not store the system in direct sunlight.
Do not use phenol-based solutions to clean the system.
Make sure the system is clean and dry prior to use or storage.
Never use sharp objects or electrically heated under blankets on or under the system.
Store the pump and mattress in the protective bags supplied.
Electromagnetic Compatibility (EMC)
This product complies with the requirements of applicable EMC Standards. Medical electrical
equipment needs special precautions regarding EMC and needs to be installed in accordance
with the following instructions:
The use of accessories not specified by the manufacturer may result in increased
emissions by, or decreased immunity of, the equipment, affecting its performance.
Portable and mobile radio frequency (RF) communications equipment (e.g. mobile/cell
phones) can affect medical electrical equipment.
If this equipment needs to be used adjacent to other electrical equipment, normal
operation must be checked before use.
For detailed EMC information contact ArjoHuntleigh service personnel.
Environmental Protection
Incorrect disposal of this equipment and its component parts, particularly batteries or other
electrical components, may produce substances that are hazardous to the environment. To
minimise these hazards, contact ArjoHuntleigh for information on correct disposal.
Service Information
ArjoHuntleigh recommend that this system should be serviced every 12 calendar months or,
where applicable, when the service indicator is illuminated.
Design Policy and Copyright
® and ™ are trademarks belonging to the ArjoHuntleigh group of companies. As our policy is
one of continuous improvement, we reserve the right to modify designs without prior notice.
© ArjoHuntleigh 2009.
1
1. Introduction
About this Manual This manual is your introduction to the Alpha Active 4.
Use it to initially set up the mattress and keep it as a
reference for day-to-day routines and as a guide to
maintenance.
About Alpha Active 4 The
Alpha Active 4 systems comprise of a mattress
replacement and pump. The support system can be used
on hospital and domestic beds.
Alpha Active 4 Pump The
Alpha Active 4 pump comprises of a moulded case
with non slip feet on the base and integral hanging
brackets.
The controls are situated on the front of the pump and a
sophisticated alarm system differentiates between
normal operation and genuine system faults. If an alarm
situation is detected an indicator will illuminate on the
front of the pump and an audible warning will sound.
Control Panel
Hanging
Mattress
Brackets
Connection
2
Alpha Active 4 Mattress Replacement
The
Alpha Active 4 mattress replacement comprises the
following components:
Detachable Cover The standard cover comprises of a 2-way stretch PU
(polyurethane) coated knitted fabric zipped to a durable
nylon base. The zips are protected by flaps to prevent
ingress of contaminants and allow easy removal of the
cover for cleaning.
Cells The mattress comprises of 19 PU cells, 16 providing
support to the user in either Alternating (Active) or
Static (Reactive) mode and 3 Static head cells.
CPR function A CPR (Cardio-Pulmonary Resuscitation) control is
positioned at the head end of the mattress to allow rapid
deflation of the mattress.
Tube-set The tube-set has a 3-way pneumatic connection which
incorporates a flexible, compact anti-kink tube that is
resistant to crushing and any subsequent obstruction of
air flow.
When disconnecting the tube-set, place the attached
cover over the end to place the mattress in transport
mode.
Mattress Replacement
Base Cover
The base cover for the mattress replacement is PVC
(Polyvinyl Chloride) coated nylon on the underside.
There are six straps located on the underside of the
mattress replacement to secure the mattress to the
bedframe.
Zipped Cover
3 Static Head Cells
16 Alternating Cells
Tubeset
Serial No.
Base Cover
(Inside cover)
CPR
3
2. Clinical Applications
Indications The Alpha Active 4 system is indicated for the
prevention and management of pressure ulcers when
combined with an individualised monitoring,
repositioning and wound care programme.
The
Alpha Active 4 mattress is designed for patients
weighting up to 200 kg (440lb).
Contra-Indications The
Alpha Active 4 system should not be used for
patients with unstable spinal fractures.
In the case of patients with other unstable fractures,
where a moving surface can be harmful, advice should
be obtained from the appropriate physician before using
the
Alpha Active 4 system.
The above are guidelines only and should not replace clinical judgement or
experience.
4
3. Installation
Preparing the Alpha Active 4 system for use
The system is very simple to set up using the following
guidelines:
1. Remove the system from the packaging. You
should have the following items:
Alpha Active 4 pump including mains power cord
and hanging brackets.
Alpha Active 4 mattress replacement with integral
tube-set.
•Cover.
Installing the Alpha Active 4 Mattress
The
Alpha Active 4 mattress replacement system should
be installed as follows:
1. Remove the existing mattress from the bed frame
and check that there are no protruding bed springs
or sharp objects on the bed frame surface.
2. Unroll the mattress onto the bed frame and ensure
that the tube-set is located near the foot end of the
bed and the CPR at the head end. The cells of the
mattress must be uppermost.
3. Secure the mattress to the bed frame using the 6
fastener straps.
If the bed can be profiled to any position (i.e. raised or lowered), attach the
mattress to the movable parts of the bed only.
To Complete the
Mattress Installation
Complete the installation of the mattress replacement as
follows:
1. If not already fitted, place the protective cover over
the mattress. Ensure that the logo is uppermost and
at the foot end of the mattress.
2. Zip the cover onto the mattress starting from the
head end and taking care not to trap any material in
the zip.
3. Ensure that the CPR unit is secured in it’s closed
position.
The CPR must be accessible at all times.
5
Installing the Pump The pump should be installed as follows:
1. Position the pump, feet down, on any convenient
horizontal surface or alternatively suspend from the
bed foot rail by means of the integral hanging
brackets.
2. Ensure that the mattress tube-set is not "kinked" or
twisted and connect it to the pump until it clicks
into place. Ensure that the tube-set is securely
connected to the pump.
3. Insert the mains power plug into a suitable mains
power socket.
System Operation The system is now ready for use. See “Operation” on
page 8 for day-to-day operating instructions.
WARNING
Make sure that the mains power cable is positioned to avoid
causing a hazard and is clear of moving bed mechanisms or other
possible entrapment areas.
6
4. Controls, Alarms and Indicators
Controls
On/Off Button Pressing the On/Off button will activate the pump.The
switch will illuminate when the pump is on.
Alarm Mute During an alarm condition the sound of the alarm can be
muted by pressing this button. The yellow indicator will
then remain on but the alarm will be muted.
Mode Select The mode selector allows the mode of operation to be
selected.
Two modes of operation are available, Alternating or
Static (non-alternating). The indicator next to the
respective icon indicates the currently selected mode.
Auto-firm Mode Selects the Auto-firm, non-alternating mode. The orange
indicator will illuminate when the pump is in this mode.
Auto-firm lasts 30 minutes, after which the pump will
default back to the previous setting.
Patient Weight Selector Two buttons (+ and -) are used to select the patient
weight. This sets the pressure inside the mattress
replacement and therefore the amount of support the
patient receives.
The pressure setting is indicated by the green indicator
above the selected setting.
Cycle Time A single button allows the cycle time to be selected for
patient and/or clinician preference.
The cycle time setting is indicated by the green indicator
above the selected setting.
AlarmsPatient Weight Selector
Cycle Time Alarm Mute Mode Select Auto-firm Mode
7
Alarms and Indicators
Low Pressure Indicator The red Low Pressure indicator is illuminated whenever
the pump detects low pressure within the mattress
replacement. An audible alarm will sound unless
cancelled by the
mute button.
The indicator will extinguish once normal pressure is
reached.
See “Troubleshooting” on page 15 for possible causes of Low Pressure.
Service Indicator/Pump Fault
The yellow Service/pump fault indicator will illuminate
and remain on if the pump has detected an internal fault.
A Service Engineer should be called.
Power Fail Indicator The red Power Fail indicator will illuminate when a
mains power failure has been detected. An audible alarm
will sound until power is resumed or the pump is
switched off using the on/off button.
8
5. Operation
These instructions cover day-to-day operation of the system. Other operations, such
as maintenance and repair, should only be carried out by suitably qualified
personnel.
Quick Start Before using the
Alpha Active 4 mattress replacement
make sure it has been installed correctly in accordance
with “Installing the Alpha Active 4 Mattress” on page 4,
and ensure that the CPR unit on the mattress is clicked
into the closed position.
1. Press the On/Off button on the side of the pump.
2. Allow approximately 30 minutes for the mattress
replacement to inflate fully.
3. Place a bed sheet loosely over the mattress without
tucking, to gain maximum pressure redistribution.
Support Setting
Procedure
It is important to follow the correct support setting
procedure to ensure the patient receives adequate
support (lift) while achieving maximum pressure
redistribution and comfort.
1. Lie or sit the patient on the mattress.
2. Select appropriate cycle time.
3. Select patient weight on pump. This should serve as
an approximate guide only. An independent clinical
determination needs to confirm that the patient is
properly supported.
4. Wait 10/20 minutes while the pump adjusts the
pressures.
5. Ensure that the patient is not ‘bottoming out’. To do
this, unfasten the cover and slide a hand beneath the
patients sacral area to ensure at least 2.5cm
clearance between the sacrum and bed base.
6. If the caregiver feels less than 2.5cm of support
material, the patient has bottomed out and the
support pressures should be adjusted accordingly.
9
Changes in Patient
Position
When a patient is in the lying or supine position, their
body weight is dispersed over a large area. When in the
sitting position, their body weight is concentrated within
a much smaller area and therefore will require more
support than in the lying position.
Therefore, when the patient changes position, it may be
necessary, in order to maximise the benefit of the
support surface, to make adjustments to the setting of the
patient weight selector.
From Lying to Sitting - Increase weight selector setting.
From Sitting to Lying - Decrease weight selector setting.
This adjustment should be in conjunction with
independent clinical determination of appropriate
support.
Static Provides a stable, non-moving support surface for
instances where active therapy is not indicated
e.g. to carry out nursing procedures or for patients
unable to tolerate a moving surface. In
Static mode the
support surface remains constant (all cells are equally
inflated). Additional nursing assessment must be
undertaken in order to direct an individualised
repositioning programme.
When operating the system in
Static mode it may be
necessary, where possible, to reduce the pressure setting
to increase patient comfort and safety.
Power Fail Condition If a Power Fail condition arises disconnect the tubeset
from the pump and place the attached cap over the end
of the tubeset to put the mattress into transport mode.
Once power is resumed, re-connect the tubeset to the
pump to continue therapy.
To Disconnect the Tube-set
To disconnect the tube-set at any time, depress the
buttons on the top and bottom of the tubeset connector
pull the tube-set connector away from the pump.
To deflate the mattress Refer to “To Deflate and Store
the Alpha Active 4 Mattress” on page 10.
10
Transport Mode To transport a patient using the Alpha Active 4 mattress
replacement, disconnect the tube-set from the pump and
place the attached cap over the end of the tubeset to put
the mattress into transport mode. This will automatically
switch the mattress into transport mode.
The patient will remain supported by the mattress for up
to 12 hours.
To resume normal operation, simply reconnect the tube-
set and run the pump.
To Deflate and Store the Alpha Active 4 Mattress
To deflate the mattress: 1. Disconnect the tube-set from the pump.
2. Activate the CPR control to deflate the mattress.
To store the mattress Following deflation:
1. Bring the tubeset over the mattress to lie parallel to
the foot end of the mattress.
2. Roll the mattress from the foot end toward the head
end of the mattress.
Caution
Transport mode is non-therapeutic offering support only for up to
12 hours. It is recommended that when in transport mode the
patient is frequently monitored.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174

GETINGE GROUP ARJOHUNTLEIGH ALPHA ACTIVE 4 Instructions For Use Manual

Typ
Instructions For Use Manual