Maintenance
Cleaning and disinfection
Clean on a regular basis. Clean surfaces with
a damp cloth using an appropriate pH-neutral
detergent. Do not use solvents or strong liquids,
this may damage surfaces on the lifter. For
disinfection when needed; use isopropyl alcohol.
Avoid abrasive cleaning products.
• Remove lifting arm cover to clean underneath
• If nessesary, remove the knee cushion and
clean it with running water
• Remove hair and pile from the castors and
verify that the castors rotates freely.
Clean contact and hand control with isopropyl
alcohol to remove grease and dirt if necessary.
Check emergency stop after cleaning.
The lift should not be exposed to running water.
Make sure not to damage or
remove labels when cleaning.
Reconditioning
Follow cleaning and assembly procedure, complete
periodic inspection and use checklist after assem-
bly to recondition the lifter.
Recycling
Refer to “Recycling instructions” for how to prop-
erly dispose of product. See www.etac.com.
Spare Parts
A list of spare parts is available on request.
Periodic Inspection Scope
Periodic inspection is a visual examination (particu-
larly of the lifter’s load bearing structure and lifting
mechanism with attachments, brakes, controls,
safety devices and person-support devices) accord-
ing to Periodic Inspection Report for Molift Quick
Raiser 205.
See www.etac.com.
If there is any problems with the
lifter that could jeopardize some-
ones safety, the lifter shall imme-
diately be taken out of service and
marked “out of order”. Do not use
the lifter untill it is repaired.
Periodic Inspection shall be performed at least
once a year or more frequently if required by local
requirements. The inspection must be performed
by service personnel authorized by Etac. Contact
Etac at molift@etac.com for training and
authorization or recommendation of an approved
service partner.
When performing a periodic inspection, the
inspector shall fill out the inspection report for
Molift Quick Raiser 205. The reports should be
retained by the person(s) responsible for servicing
the lifter. If the inspection reveals defects and
damages, the lifter shall be taken out of operation
and marked “out of order” until it is repaired.
After performing periodic inspection reset the
service indicator, see page 11
Periodic Inspection
20
Molift Smart 150 / www.etac.com