Medtronic guardian 2 link Användarmanual

Typ
Användarmanual
Transmitter
Transmetteur
Transmitter
Transmisor
Zender
Trasmettitore
Sender
Lähetin
Sändare
Sender
Transmissor
Transmissor
MP6025871-12C1 / A
RELEASED
MP6025871-12C1 / A
RELEASED
Contacts:
Africa: Medtronic Africa (Pty) Ltd.
Tel: +27 (0) 11 677 4800
Argentina: Corpomedica S.A.
Tel: +(11) 4 814 1333
Medtronic Directo 24/7: +0800 333 0752
Australia: Medtronic Australasia Pty. Ltd.
Tel: 1800 668 670 (product orders)
Tel: 1800 777 808 (customer help)
Azerbaijan: Isomed
Tel: +994 (12) 464 11 30
Bangladesh: Sonargaon Healthcare Pvt Ltd.
Mobile: (+91)-9903995417
or (+880)-1714217131
Belarus: ОДОБаджин
Tel: +375 17 313 0990
België/Belgique: N.V. Medtronic Belgium S.A.
Tel: 0800-90805
Bosnia and Herzegovina: Medimpex d.o.o.
Tel: +387 33 476 444
or +387 33 476 400
Fax: +387 33 476 401
or +387 33 432 241
Brasil: Medtronic Comercial Ltda.
Tel: +(11) 2182-9200
Medtronic Directo 24/7: +0800 773 9200
Bulgaria: RSR Ltd.
Tel: +359 885 428 900
Canada: Medtronic of Canada Ltd.
Tel: 1-800-284-4416 (toll free/sans-frais)
Chile: Medtronic Chile
Tel: +(9) 66 29 7126
Medtronic Directo 24/7: +1 230 020 9750
Medtronic Directo 24/7 (From Santiago): +(2)
595 2942
China: Medtronic (Shanghai) Ltd.
24 Hour Help (Cell): +86 400-820-1981
24 Hour Help (Landline): +86 800-820-1981
Colombia: Medtronic Latin America Inc. Sucursal
Colombia
Tel: +(1) 742 7300
Medtronic Directo 24/7 (Landline): +01 800 710
2170
Medtronic Directo 24/7 (Cellular): +1 381 4902
Croatia: Medtronic Adriatic d.o.o.
Tel: +385 1 488 11 20
Fax: +385 1 484 40 60
Danmark: Medtronic Danmark A/S
Tel: +45 32 48 18 00
Deutschland: Medtronic GmbH
Geschäftsbereich Diabetes
Telefon: +49 2159 8149-370
Telefax: +49 2159 8149-110
24-Stdn-Hotline: 0800 6464633
Eire: Accu-Science LTD.
Tel: +353 45 433000
España: Medtronic Ibérica S.A.
Tel: +34 91 625 05 42
Fax: +34 91 625 03 90
24 horas: +34 900 120 330
Europe: Medtronic Europe S.A. Europe, Middle
East and Africa Headquarters
Tel: +41 (0) 21-802-7000
France: Medtronic France S.A.S.
Tel: +33 (0) 1 55 38 17 00
Hellas: Medtronic Hellas S.A.
Tel: +30 210677-9099
Hong Kong: Medtronic International Ltd.
Tel: +852 2919-1300
To order supplies: +852 2919-1322
24-hour helpline: +852 2919-6441
India: I
ndia Medtronic Pvt. Ltd
Tel: (+91)-
80-22112245 / 32972359
Mobile: (+91)-9611633007
Indonesia: Medtronic International Ltd.
Tel: +65 6436 5090
or +65 6436 5000
MP6025871-12C1 / A
RELEASED
Israel: Agentek
Tel: +972 3649 3111
Italia: Medtronic Italia S.p.A.
Tel: +39 02 24137 261
Fax: +39 02 24138 210
Servizio assistenza tecnica:
Nº verde 24h: 800 20 90 20
Japan: Medtronic Japan Co. Ltd.
Tel: +81-3-6430-2019
24 Hr. Support Line: 0120-56-32-56
Kazakhstan: Medtronic BV in Kazakhstan
Tel: +7 727 311 05 80 (Almaty)
Tel: +7 717 224 48 11 (Astana)
 Круглосуточная линия поддержки: 8 800 080
5001
Latin America: Medtronic, Inc.
Tel: 1(305) 500-9328
Fax: 1(786) 709-4244
Latvija: Ravemma Ltd.
Tel: +371 7273780
Macedonia: Alkaloid Kons Dooel
Tel: +389 2 3204 430
Magyarország: Medtronic Hungária Kft.
Tel: +36 1 889 0688
Malaysia: Medtronic International Ltd.
Tel: +603 7946 9000
Middle East and North Africa: Regional Office
Tel: +961-1-370 670
Montenegro: Glosarij
Tel: +382 20 642 495
Fax: +382 20 642 540
México: Medtronic Servicios S. de R. L. de C.V.
Tel (México DF): +(11) 029 058
Tel (Interior): +01 800 000 7867
Medtronic Directo 24/7 (from México DF):
+(55) 36 869 787
Medtronic Directo 24/7: +01 800 681 1845
Nederland, Luxembourg: Medtronic B.V.
Tel: +31 (0) 45-566-8291
Gratis: 0800-3422338
New Zealand: Medica Pacifica
Phone: 64 9 414 0318
Free Phone: 0800 106 100
Norge: Medtronic Norge A/S
Tel: +47 67 10 32 00
Fax: +47 67 10 32 10
Philippines: Medtronic International Ltd.
Tel: +65 6436 5090
or +65 6436 5000
Poccия: ООО «Медтроник»
Tel: +7 495 580 73 77
 Круглосуточная линия поддержки: 8 800 200 76
36
Polska: Medtronic Poland Sp. Z.o.o.
Tel: +48 22 465 6934
Portugal: Medtronic Portugal Lda
Tel: +351 21 7245100
Fax: +351 21 7245199
Puerto Rico: Medtronic Puerto Rico
Tel: 787-753-5270
Republic of Korea: Medtronic Korea, Co., Ltd.
Tel: +82.2.3404.3600
Ro
man
ia: Medtronic BV Reprezentanta
Tel: +40 372 188 000
Schweiz: Medtronic (Schweiz) AG
Tel: +41 (0)31 868 0160
24-Stunden-Hotline: 0800 633333
Fax Allgemein: +41 (0)318680199
Serbia: Medtronic B.V. Serbia
Tel: +381 11 2095 900
Singapore: Medtronic International Ltd.
Tel: +65 6436 5090
or +65 6436 5000
Slovenija: Zaloker & Zaloker d.o.o.
brezplačna številka: 080 1880
Tel: +386 1 542 51 11
Slovenská republika: Medtronic Slovakia, s.r.o.
Tel: +421 26820 6942
HelpLine: +421 26820 6986
MP6025871-12C1 / A
RELEASED
Sri Lanka: Swiss Biogenics Ltd.
Mobile: (+91)-9003077499
or (+94)-777256760
Suomi: Medtronic Finland Oy
Tel: +358 20 7281 200
Help line: +358 400 100 313
Sverige: Medtronic AB
Tel: +46 8 568 585 20
Fax: +46 8 568 585 11
Taiwan: Medtronic (Taiwan) Ltd.
Tel: 02-21836000
Toll free: +886-800-005285
Thailand: Medtronic (Thailand) Ltd.
Tel: +662 232 7400
Türkiye: Medtronic Medikal Teknoloji
Ticaret Ltd. Sirketi.
Tel: +90 216 4694330
USA: Medtronic Diabetes Global Headquarters
24 Hour HelpLine: +1-800-646-4633
To order supplies: +1-800-843-6687
Ukraine: Medtronic B.V. Representative office in
Ukraine
Tel: +38 044 392 04 01
 Лінія цілодобової підтримки:
0 800 508 300
United Kingdom: Medtronic Ltd.
Tel: +44 1923-205167
Österreich: Medtronic Österreich GmbH
Tel: +43 (0) 1 240 44-0
24 – Stunden – Hotline: 0820 820 190
Česká republika: Medtronic Czechia s.r.o.
Tel: +420 233 059 401
Non-stop helpLine:
+420 233 059 059
MP6025871-12C1 / A
RELEASED
MP6025871-12C1 / A
RELEASED
The Guardian™ 2 Link transmitter is a component of the continuous glucose
monitoring system for the MiniMed
640G insulin pump. The transmitter collects data
from the Enlite™ glucose sensor. The transmitter then wirelessly sends the data to
the insulin pump.
Guardian 2 Link transmitter kit components
A complete Guardian 2 Link transmitter kit includes the following components:
  Guardian 2 Link transmitter (MMT-7731)
  Watertight Tester (MMT-7726)
  Charger (MMT-7715)
  One-press serter (MMT-7512)
Indications for use
The transmitter is indicated for single-patient use as a component of select Medtronic
continuous glucose sensing systems and MiniMed sensor-enabled pump systems.
Contraindications
Do not expose your transmitter to MRI equipment, diathermy devices, or other
devices that generate strong magnetic fields. If your transmitter is inadvertently
exposed to a strong magnetic field, discontinue use and contact the 24 Hour
HelpLine or your local representative for further assistance.
Warnings
This product contains small parts and may pose a choking hazard for young children.
If the tester comes in contact with blood, the tester must be discarded. Dispose of the
tester according to the local regulations for medical waste disposal.
Bleeding may occur after inserting the sensor. Make sure that the site is not bleeding
before connecting the transmitter to the sensor. If bleeding occurs, apply steady
pressure with a sterile gauze or clean cloth at the insertion site until bleeding stops.
After bleeding stops, connect the transmitter to the sensor.
Contact the 24 Hour HelpLine or your local representative if you experience any
adverse reactions associated with the transmitter or sensor.
-1-
English
MP6025871-12C1 / A
RELEASED
Magnetic fields
Do not expose your transmitter to MRI equipment, diathermy devices, or other devices
that generate strong magnetic fields. If your transmitter is inadvertently exposed to a
strong magnetic field, discontinue use and contact the 24 Hour HelpLine or your local
representative for further assistance.
X-rays, MRIs, diathermy devices, and CT scans
If you are going to have an x-ray, diathermy treatment, CT scan, MRI or other type of
exposure to radiation, remove your sensor and transmitter before entering a room
containing any of these equipment.
Important information about airport security systems, and using your transmitter on an
airplane, can be found on the Emergency Card. Be sure to carry the Emergency Card
provided with your device when you are traveling.
Precautions
Refer to the Enlite Sensor User Guide for all precautions, warnings, and instructions
relating to the sensor.
Always use the tester when cleaning the transmitter. Do not use any other test plug with
the transmitter.
Do not twist the tester or sensor while attached to the transmitter. This will damage the
transmitter.
Do not allow water, or any other liquid, to come in contact with the tester when it is not
connected to the transmitter. A wet tester can cause damage to the transmitter.
Do not allow the transmitter to come in contact with any liquid when not connected to a
sensor or to the tester.
Do not clean the o-rings on the tester, as this can damage the o-rings.
Notice
Caution: Any changes or modifications to the devices not expressly approved by
Medtronic Diabetes could interfere with your ability to operate the
equipment, cause injury, and void your warranty.
Radio Frequency (RF) communication
This device complies with the United States Federal Communications Commission (FCC)
and international standards for electromagnetic compatibility.
-2-
MP6025871-12C1 / A
RELEASED
This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions: (1) This device may not cause harmful interference, and (2)
this device must accept any interference received, including interference that may
cause undesired operation.
These standards are designed to provide reasonable protection against excessive
radio frequency interference, and prevent undesirable operation of the devices from
unwanted electromagnetic interference.
This equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to Part 15 of the FCC rules. These limits are designed to
provide reasonable protection against harmful interference in a residential
installation. This equipment generates, uses, and can radiate radio frequency energy
and, if not installed and used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
  Reorient or relocate the receiving antenna.
  Increase the separation between the equipment and the receiver.
This device can generate, use, and radiate radio frequency energy and, if installed
and used in accordance with the instructions, may cause harmful interference to
radio communications. If the device does cause interference to radio or television
reception, you are encouraged to try to correct the interference by one or more of the
following measures:
  Decrease the distance between the transmitter and the insulin pump to 1.8
meters (6 feet) or less.
  Increase the separation between the transmitter and the device that is receiving/
emitting interference.
If other devices that employ radio frequencies are in use, such as cell phones,
cordless phones, and wireless networks, they may prevent communication between
the transmitter and the insulin pump. This interference does not cause any incorrect
data to be sent and does not cause any harm to your devices. Moving away from, or
turning off, these other devices may enable communication. If you continue to
experience RF interference, please contact your local representative.
Caution: Changes or modifications to the internal RF transmitter or antenna
not expressly approved by Medtronic could void the user's
authority to operate this insulin delivery system.
-3-
English
MP6025871-12C1 / A
RELEASED
For Canada only
This device complies with Industry Canada licence-exempt RSS standard(s).
Operation is subject to the following two conditions: (1) this device may not cause
interference, and (2) this device must accept any interference, including
interference that may cause undesired operation of the device.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux
appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions
suivantes : (1) l'appareil ne doit pas produire de brouillage, et (2) l'appareil doit
accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible
d'en compromettre le fonctionnement.
Directive 1999/5/EC
Medtronic declares that this product is in conformity with the essential requirements of
Directive 1999/5/EC on Radio and Telecommunications Terminal Equipment.
For additional information, contact Medtronic MiniMed at the address or phone number
provided on the back cover.
Assistance
Please contact your local representative for assistance. Refer to the Medtronic Diabetes
International Contacts list at the beginning of this user guide for contact information.
Charger
The transmitter contains a non-replaceable, rechargeable battery that you can recharge
as needed with the charger. The charger has a green light that shows the charging status
and a red light that communicates any problems during charging. If you see a red light,
see the Troubleshooting section. The charger needs one AAA alkaline battery.
Note: If the battery is installed incorrectly or is low, the charger will not work. Repeat the
battery installation steps using a new battery.
Installing a battery in the charger
To install a battery in the charger:
1
Push the battery cover in and slide it off (as shown in the following illustration).
2
Insert a new alkaline AAA battery. Make sure the + and - symbols on the battery align
with these same symbols shown on the charger.
-4-
MP6025871-12C1 / A
RELEASED
3Slide the cover back on the charger until it clicks into place.
Charging the transmitter
Caution: Charge the transmitter after each sensor use. A fully charged
transmitter works at least six days without recharging. A depleted
transmitter can take up to one hour to recharge.
To charge the transmitter:
1
Connect the transmitter to the charger by lining it up, flat
side down, with the charger. Push the two components
together fully.
2
Within 10 seconds after the transmitter is connected, a
green light on the charger will flash for one to two seconds
as the charger powers on. For the rest of the charging time,
the charger’s green light will continue to flash in a pattern of
four flashes with a pause between the four flashes.
3
When charging is complete, the green light on the charger
will stay on, without flashing, for 15 to 20 seconds and then
turn off.
4
After the green charger light turns off, disconnect the
transmitter from the charger. The green light on the
transmitter will flash for about five seconds and then turn
off.
Inserting the sensor
Always refer to the serter user guide for instructions on how to insert the sensor.
-5-
English
MP6025871-12C1 / A
RELEASED
Connecting the transmitter to the sensor
Before proceeding, have your pump user guide available.
To connect the transmitter to the sensor:
1
After the sensor is inserted, consult your serter user guide for details on applying the
required overtape.
2
Hold the rounded end of the inserted sensor to prevent it from moving during
connection.
3
Hold the transmitter as shown. Line up the two notches on the
transmitter with the side arms of the sensor. The flat side of the
transmitter should face the skin.
4
Slide the transmitter onto the sensor until the sensor’s flexible
arms snap into the notches on the transmitter. If the transmitter
is properly connected, and if the sensor has had enough time to
become hydrated, the green light on the transmitter will flash
within 10 seconds.
5
If the transmitter light does not flash, disconnect the transmitter from the sensor, wait
for several seconds and then reconnect. If the transmitter light still does not flash,
charge the transmitter.
6
When the transmitter light flashes green when connected to the sensor, use your
pump to start the sensor. For more instructions, see your pump user guide.
7
After the transmitter successfully sends sensor data to the pump, attach the sensor's
adhesive tab to the transmitter.
8
Follow the instructions that appear on the pump screen or follow the instructions in
your pump user guide.
-6-
MP6025871-12C1 / A
RELEASED
Disconnecting the transmitter from the sensor
Before proceeding, have your pump user guide available.
To disconnect the transmitter from the sensor:
1
Carefully remove any occlusive dressing from the
transmitter and sensor.
2
For the Enlite sensor, remove the adhesive tab from the top
of the transmitter.
3
Hold the transmitter as shown, and pinch the flexible side
arms of the sensor between your thumb and forefinger.
4
Gently pull the transmitter away from the sensor.
5
Follow the instructions that appear on the pump or follow
the instructions in your pump user guide.
Removing the sensor
Always refer to the sensor user guide for instructions on how to remove the sensor.
Bathing and swimming
After the transmitter and sensor are connected, they form a waterproof seal to a
depth of 2.4 meters (eight feet) for up to 30 minutes. You can shower and swim
without removing them. No occlusive dressing or overtape is needed.
Watertight Tester
The tester is used to test the transmitter to make sure it is working. It is also used as
a required component for cleaning the transmitter. Properly connecting the tester to
the transmitter will ensure that fluids do not come in contact with the transmitter’s
connector pins. Fluids can cause connector pins to corrode and affect the
transmitter’s performance.
Do not twist the tester while attached to the transmitter. This will damage the
transmitter.
The tester can be used for one year. If you continue to use the tester for more than
one year, the transmitter's connector pins could be damaged, because the tester
cannot continue to provide a waterproof seal. For instructions on how to check the
connector pins, see Inspecting the transmitter connector pins, on page 8.
-7-
English
MP6025871-12C1 / A
RELEASED
Caution: Only use the tester with the transmitter. Do not use any other test plug.
i
Inspecting the transmitter connector pins
This image is an example of how the connector pins should look.
connector opening
connector pins
housing
Look inside the transmitter’s connector opening to make sure that the connector pins are
not damaged or corroded. If the connector pins are damaged or corroded, the transmitter
cannot communicate with the charger or pump. Contact the 24 Hour HelpLine or your
local representative. It may be time to replace your transmitter.
Also look for moisture inside the connector opening. If you see any moisture, allow the
transmitter to dry for at least one hour. Moisture inside the connector opening could
cause the transmitter to not work properly, and could cause corrosion and damage over
time.
-8-
MP6025871-12C1 / A
RELEASED
Connecting the tester for testing or cleaning
Before proceeding, have available your pump user guide.
To connect the tester:
1
Hold the transmitter and the tester as shown. Line up the
flat side of the tester with the flat side of the transmitter.
2
Push the tester into the transmitter until the flexible side
arms of the tester click into the notches on both sides of the
transmitter.
3
Within five seconds, when properly connected, the green
light on the transmitter flashes for about 10 seconds.
4
To test the transmitter, check the sensor icon on the pump
to ensure that the transmitter is sending a signal (see your pump user guide).
5
To clean the transmitter, see Cleaning the transmitter, on page 9.
6
After testing or cleaning, disconnect the tester from the transmitter.
Disconnecting the tester
To disconnect the tester:
1
Hold the transmitter body as shown and pinch the side
arms of the tester.
2
With the tester arms pinched, gently pull the transmitter
away from the tester.
Note: To save transmitter battery life, do NOT leave the
tester connected after cleaning or testing.
Cleaning the transmitter
The transmitter is a single-patient use device and not intended for multi-patient use.
Caution: Do not discard the transmitter in a medical waste container or
otherwise subject it to incineration. The transmitter contains a
battery that may explode upon incineration.
Note: The tester is a required component for cleaning the transmitter. For details,
see Watertight Tester, on page 7.
Always clean the transmitter after each use.
To clean the transmitter, you will need the following materials: mild liquid soap, a soft-
bristled toddler toothbrush, a container, 70% isopropyl alcohol, and a few clean, dry
cloths.
-9-
English
MP6025871-12C1 / A
RELEASED
Warning: Cracking, flaking, or damage of the housing are signs of deterioration
and the performance of the device may be compromised. This may
affect the ability to properly clean and disinfect the transmitter. If these
signs are noted, stop using the device and call the 24 Hour HelpLine or
your local representative. The device must be discarded according to
local regulations for battery disposal (non-incineration).
To clean the transmitter:
1
Attach the tester to the transmitter.
2
If optional occlusive dressing or overtape was used and there is adhesive residue on
the transmitter, see Removing adhesive residue, on page 10.
3
Rinse the transmitter under room temperature tap water for at least one minute, and
until visibly clean. Make sure all hard-to-reach areas are rinsed completely.
4
Prepare a mild liquid soap solution using 5 milliliters (1 teaspoon) of liquid soap per
3.8 liters (1 gallon) of room temperature tap water.
5
With tester still attached, submerge the transmitter in the cleaning solution and soak
for one minute.
6
Holding the tester, brush the entire surface of the transmitter using a soft-bristled
toddler toothbrush. Make sure to brush all hard-to-reach areas until visibly clean.
7
Rinse the transmitter under running room temperature tap water for at least one
minute, and until all visible liquid soap is gone.
8
Dry the transmitter and tester with a clean, dry cloth.
9
Place the transmitter and tester on a clean, dry cloth and air dry them completely.
10
Disconnect the tester from the transmitter.
Removing adhesive residue
You may need to perform this procedure only if you have used optional occlusive
dressing, which may leave adhesive residue on the transmitter. If you visually inspect the
transmitter and see adhesive residue on it, follow the instructions below.
To remove adhesive residue, you will need the following materials: medical adhesive
remover (such as Detachol
®
- which is a mineral spirit) and cotton swabs.
Note: During testing, Medtronic MiniMed used Detachol to remove the adhesive residue
from the transmitter. Detachol is recommended for use but may not be available in
all countries.
To remove adhesive residue:
1
Make sure the tester is attached to the transmitter.
-10-
MP6025871-12C1 / A
RELEASED
2Holding the tester, saturate a cotton swab in the Detachol solution and gently rub
the adhesive residue on the transmitter until it is fully removed.
3
Continue with the cleaning procedure. See Cleaning the transmitter, on page 9
for details.
Cleaning the charger
This procedure is for general cleaning as required, based on physical appearance.
Caution: The charger is NOT waterproof. Do NOT immerse in water or any
other cleaning agent.
Caution: Dispose the charger according to the local regulations for battery
disposal (non-incineration).
To clean the charger:
1
Wash your hands thoroughly.
2
Use a damp cloth with mild cleaning solution, such as a dishwashing detergent,
to clean any dirt or foreign material from the outside of the charger. Never use
organic solvents, such as paint thinner or acetone, to clean the charger.
3
Place the charger on a clean, dry cloth and air dry for two to three minutes.
Troubleshooting
The following table contains troubleshooting information for the transmitter, charger,
and tester. For more information about troubleshooting, see your pump user guide.
Problem Likely Cause(s) Resolution
You connected the
transmitter to the
charger and no lights
came on.
The transmitter connec-
tor pins are damaged or
corroded.
Your charger battery has
no power.
1Check the transmitter connector pins for
damage or corrosion. For more informa-
tion about your connector pins, see In-
specting the transmitter connector pins,
on page 8. If the pins are damaged or cor-
roded, contact the 24 Hour HelpLine or
your local representative. It may be time
to replace your transmitter.
2
If there is no damage to the connector
pins, replace the battery in the charger.
For instructions on replacing your charger
battery, see Installing a battery in the
charger, on page 4.
-11-
English
MP6025871-12C1 / A
RELEASED
Problem Likely Cause(s) Resolution
During charging, the
flashing green light on
the charger turns off
and you see a flashing
red light on the charg-
er.
Your charger battery is
low on power.
Replace the battery in the charger. For instruc-
tions on replacing your charger battery, see In-
stalling a battery in the charger, on page 4.
During charging, the
flashing green light on
the charger turns off
and you see a series of
quick flashing red lights
on the charger.
Your transmitter is low
on power.
1
Charge the transmitter continuously for
one hour. If flashing does not stop, pro-
ceed to step 2.
2
Charge the transmitter continuously for
eight hours. If flashing does not stop, call
the 24 Hour HelpLine or your local repre-
sentative. It may be time to replace your
transmitter.
During charging, a mix
of quick and long flash-
ing red lights appear
on the charger.
Your charger and your
transmitter are low on
power.
1
Replace the battery in the charger. For in-
structions on replacing your charger bat-
tery, see Installing a battery in the
charger, on page 4.
2
Charge the transmitter continuously for
one hour. If flashing does not stop, pro-
ceed to step 3.
3
Charge the transmitter continuously for
eight hours. If flashing does not stop, call
the 24 Hour HelpLine or your local repre-
sentative. It may be time to replace your
transmitter.
The green light on the
transmitter does not
flash when you con-
nect it to the sensor.
Your transmitter is not
fully connected.
Your transmitter is low
on power.
Your sensor is not prop-
erly inserted into your
body.
1
Disconnect the transmitter from the sen-
sor.
2
Wait for five seconds and reconnect them.
If the green light still does not flash, pro-
ceed to step 3.
3
Fully charge the transmitter. If the green
light still does not flash, proceed to step 4.
4
The sensor may not be properly inserted
into your body. Insert a new sensor.
-12-
MP6025871-12C1 / A
RELEASED
Problem Likely Cause(s) Resolution
The green light on the
transmitter does not
flash when you con-
nect it to the tester.
Your transmitter is low
on power.
Your transmitter is not
fully connected.
1Check the connection between the trans-
mitter and the tester. If the green light still
does not flash, proceed to step 2.
2
Fully charge the transmitter.
3
Test the transmitter with the tester again.
If you still do not see the green light flash,
call the 24 Hour HelpLine or your local
representative. It may be time to replace
your transmitter.
Your transmitter bat-
tery does not last for
six days.
Your transmitter is not
fully charged when you
connect it to the sensor.
The transmitter and
pump frequently lose
wireless connection.
1
Fully charge the transmitter before con-
necting it to the sensor. If the transmitter
battery still does not last for six days, pro-
ceed to step 2.
2
Move away from any device that can
cause RF interference. For more informa-
tion on RF interference, see Radio Fre-
quency (RF) communication, on page 2.
3
Make sure your pump and your transmitter
are located on the same side of your body
to minimize any RF interference. If your
fully charged transmitter battery continues
to lose power before a full six days, call
the 24 Hour HelpLine or your local repre-
sentative. It may be time to replace your
transmitter.
Your transmitter has
lost connection with
your pump.
Your pump is out of
range.
There is RF interference
from other devices.
1
Move away from any device that can
cause RF interference. For more informa-
tion on RF interference, see Radio Fre-
quency (RF) communication, on page 2. If
your transmitter is still not communicating
with your pump, proceed to step 2.
2
Make sure your pump and your transmitter
are located on the same side of your body
to minimize any RF interference. If your
transmitter is still not communicating with
your pump, call the 24 Hour HelpLine or
your local representative for assistance.
-13-
English
MP6025871-12C1 / A
RELEASED
Storing the devices
Store the transmitter, charger, and tester in a clean, dry location at room temperature. If
the transmitter is not in use, you must charge the transmitter at least once every 60 days.
Although not required, you may store the transmitter on the charger. If you are storing the
transmitter on the charger, you must disconnect and reconnect the charger and the
transmitter at least once every 60 days.
Disposal
Discard the transmitter according to local regulations for battery disposal.
Specifications
Biocompatibility Transmitter: Complies with EN ISO 10993-1
Applied parts Transmitter
Sensor
Operating conditions Transmitter temperature: -5 °C to 45 °C (23 °F to 113 °F)
Caution: When operating the transmitter on a tester in air temperatures
greater than 41 °C (106 °F), the temperature of the transmitter may exceed
43 °C (109 °F)
Transmitter relative humidity: 5% to 95% with no condensation
Transmitter pressure: 61.36 kPa to 106.17 kPa (8.9 psi to 15.4 psi)
Charger temperature: 10 °C to 40 °C (50 °F to 104 °F)
Charger relative humidity: 30% to 75% with no condensation
Storage conditions Transmitter temperature: -25 °C to 55 °C (-13 °F to 131 °F)
Transmitter relative humidity: 10% to 100% with condensation
Transmitter pressure: 61.36 kPa to 106.17 kPa (8.9 psi to 15.4 psi)
Charger temperature: -10 °C to 50 °C (14 °F to 122 °F)
Charger relative humidity: 10% to 95% with no condensation
Battery life Uses one new AAA battery to charge the transmitter.
Transmitter frequen-
cy
2.4 GHz, 2M65G1D modulation, less than 1mW ERP
Maximum output
power (EIRP)
-0.63 dBm
-14-
MP6025871-12C1 / A
RELEASED
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268
  • Page 269 269
  • Page 270 270
  • Page 271 271
  • Page 272 272
  • Page 273 273
  • Page 274 274
  • Page 275 275
  • Page 276 276
  • Page 277 277
  • Page 278 278
  • Page 279 279
  • Page 280 280
  • Page 281 281
  • Page 282 282

Medtronic guardian 2 link Användarmanual

Typ
Användarmanual